To save battery power, leave Wi-Fi disabled during everyday use and enable it only when needed. If a solenoid valve has shut off the water after an alarm, use the Resume water function.
Before you start
On-demand Wi-Fi and solenoid valve.
Use the Resume water function from the screen or web interface if a solenoid valve is used.
Quick start
Basic checks during production.
Switch on CP-FLOW
Leave the display on the Flow page while producing water.
Open the water supply
The flow should change from 0.00 to a measured value.
Wake the display if needed
The display may turn off after about 1 minute. Turn or press the knob to wake it.
Using the rotary knob
Local CP-FLOW navigation.
Wi-Fi and web dashboard
Enable Wi-Fi only when needed.
Enable Wi-Fi: ON
From the CP-FLOW menu.
Connect to the network
Choose CP-FLOW-xxxx with password 123456789.
Open the interface
Enter http://192.168.4.1. On iPhone, make sure you type http:// before the address.
Counters, cartridges and alarms
Produced volume and maintenance management.
Flow and session
Instant flow is shown in L/min and the session tracks the volume produced during the current run.
Cartridges
CP-FLOW can manage up to 4 cartridges with name, capacity, threshold and alarm.
Resetting counters
Use the Reset command for the cartridge if you replace only one element, or Reset pack if you replace the entire pack.
Check measurement accuracy
Verify the measurement with a known volume before making any correction.
Prepare exactly 1 litre
Use a reliable graduated container and measure a true volume of 1 litre.
Compare with CP-FLOW
Run this litre through the sensor and note the value shown by CP-FLOW.
If a difference remains, contact support
The correction method depends on the software version. Send support the actual value, the displayed value and a photo or screenshot of the settings screen; CoralPlast will tell you which field and value to use for your version.
Logs and OTA update
Maintenance functions.
Logs
Enable Wi-Fi and open the interface to view available sessions and diagnostics.
OTA
The update page is available at http://192.168.4.1/update when Wi-Fi is enabled.
Quick troubleshooting
Simple checks before contacting support.
The web page does not open
Check that Wi-Fi: ON is enabled, that you are connected to CP-FLOW-xxxx and that you enter http://192.168.4.1.
The water remains shut off
Use Resume water from the display or web dashboard.
The measurement looks inaccurate
Repeat the accuracy check using exactly 1 litre measured in a reliable container.
Check
- Flow visible during production.
- Rotary knob tested.
- Cartridges configured.
- Calibration/check performed if necessary.
Avoid
- Leaving Wi-Fi enabled unnecessarily if you want to save battery power.
- Forgetting Resume water after a solenoid shut-off.
- Calibrating with an approximate volume.
Need personalised help?
Tell us the exact model, describe the behaviour you observe and, if possible, attach a photo or short video of the installation.